Bosch Rexroth Canada ı 16.10.2008 ı Revision 2.2
The resulting “Weight per minute” calibration measurement can be seen in P23
P23P23
P23
(open loop control).
The resulting “Weight per revolution” calibration measurement can be seen in P24
P24P24
P24
(closed loop control).
If verification of the “Material calibration” indicates that the calibration is
inaccurate, an adjustment can be made using P23
P23P23
P23 or P24
P24P24
P24. If the discharge weight is
lower than expected set P23/
P23/P23/
P23/P24
P24P24
P24 lower, if the discharge weight is higher than
expected set P23/
P23/P23/
P23/P24
P24P24
P24 higher.
5.5 Liquid Calibration
Fill the liquid reservoir with the liquid to be calibrated
Operate the vehicle engine at 150 RPM
Set the controller to P27
P27P27
P27 and Press the Conveyor Dial
Press the UPPER SELECT key to start the calibration process
Allow the liquid to discharge for exactly one (1) minute
Press the UPPER SELECT key to stop the liquid discharge
Measure the liquid discharged during the calibration process and enter the
volume using the SELECT keys
Press BLAST to accept this value
If verification of the “Material calibration” indicates that the calibration is
inaccurate, an adjustment can be made using P28
P28P28
P28. If the discharge weight is lower
than expected set P28
P28P28
P28 lower, if the discharge weight is higher than expected set
P28
P28P28
P28 higher.