Bosch NurseCall N46 User Manual Page 20

  • Download
  • Add to my manuals
  • Print
  • Page
    / 36
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 19
20 en | Use NurseCall N46 Wall Transmitter
970.030 | v1.0 | 2009.06 User Manual Bosch Security Systems
6Use
Every time a NurseCall N46 Wall Transmitter is used by pressing a button or activating a wired
input, the device sends a radio signal and the lamps indicates its actual state. Optionally, two
relays allow the transmission of status to a remote display such as the signal lamps placed on
top of the door.
Each radio message is sent with a different code which allows the receiver (NurseCall system)
to recognize the performed action.
6.1 Call for Help
The Call for Help can be triggered in three different ways:
by pushing the red Call for Help button
by pushing the external switch connected to the J4 connector, for example a pear
push button.
by pushing the external button connected to terminals 1 & 2 of the terminal block
J6.
The Call for Help is repeated approximately every 2 minutes until a call acknowledgement
takes place, up to a maximum of 20 minutes. This repeat function can be disabled by setting
the solder bridge B4.
A Call for Help is signaled by red flashes on the signal lamp and optionally, by the activation of
the relay Call for Help.
6.2 Call for Assistance
The Call for Assistance is triggered by pushing the Call for Assistance button (blue).
The Call for Assistance is repeated approximately every 2 minutes until a call acknowledge-
ment takes place, up to a maximum of 20 minutes. This repeat function can be disabled by
setting the solder bridge B4.
A Call for Assistance is signaled by red double flashes on the signal lamp and by the alternate
action/release of the relay Call for Help.
i
NOTICE!
It is recommended to keep the button pressed until the lamp placed above the red button
lights up.
i
NOTICE!
In some cases, the action Call for Help generates a Call for Assistance action. See
Section State before action: Alarm with staff presence, page 25.